TTP-VARIO SYSTEM
Patented clamping mechanism directly within the head plate allows for shortest Functional Length of 0.75 mm.
Conically shaped bell: optimally adapts to the stapes head encouraging acoustic transfer function in the direction of the oval window. Recessed slots allow adequate space for the stapedius tendon and are easily augmented in the in case of superstructure anomaly.

TTP-VARIO Partial Prosthesis

Adjustable lenght prosthesis

 

  • Fine, elegant design
  • Fenestraded head plate
  • Partially roughened surface
  • Pure titanium
  • Finely balanced weight distribution
  • MR conditional for the future: 7.0 Tesla

TTP-VARIO Total Prosthesis

Adjustable lenght prosthesis

 

  • Fine, elegant design
  • Fenestrated head plate
  • Partially roughened surface
  • Pure titanium
  • Finely balanced weight distribution
  • MR conditional for the future: 7.0 Tesla
  • The TTP-VARIO Total Prosthesis is compatible with the OMEGA CONNECTOR
Patented clamping mechanism directly within the headplate allows for shortest length of 3.0 mm
Cannulated distal footing developed to give the prosthesis an excellent degree of stability

Product Information

PRECISION LENGTH ADJUSTMENT

 

TTP-VARIO Titanium Prostheses provide the unique benefits of a length adjustable prosthesis system. The patented clamping mechanism within the head plate securely fixates the head plate to the stem. It provides a flexible design enabling prostheses to be sized as short as 0.75 mm (FL). This system satisfies all lengths within two types of prostheses: one for partial reconstructions and one for total reconstructions. Such flexible use makes a significant contribution to optimizing inventory levels.

Precise length adjustments are made with the aid of the TTP-VARIO instrument set.

Developed in close collaboration with Tuebingen University ENT Clinic, Germany.

GOOD TO KNOW

The tension-sensitive annular ligament of the stapes footplate plays a key role in reconstruction of the ossicular chain: If undesirable pretension develops here, it can have a critical effect on postoperative hearing outcome. Short prosthetic solutions as short as 0.75 mm Functional Length are available to minimize risk of unintentional tension.

See also:

Voss, S. E., Nakajima, H. H., Huber, A. M., & Shera, A. C. (2013). Function and Acoustics of the Normal and Diseased Middle Ear. In Puria, S., Fay, R. R. & Popper, A. (Eds). The Middle Ear. Science, Otosurgery, and Technology. Chapter 4. New York, Heidelberg, Dordrecht, London: Springer.

Benefits

  • Highly flexible system to ensure most appropriate length and provides maximum variability
  • Major contribution to the optimization of inventory levels
  • Maximum intraoperative visibility
  • Improved stability and handling
  • Clinically evaluated
  • Excellent biocompatibility

Ordering Information

Ordering Information - TTP-VARIO BELL Partial Prosthesis
ITEM REF
TTP-VARIO BELL Partial Prosthesis Adjustable Length
1.75 - 4.50 mm (in 0.25 increments)
Functional Length (FL) 0.75 - 3.50 mm
1002 010

 

Material: Pure Titanium (ASTM F67 Medical Grade)

Shaft Diameter: 0.2 mm

 

Ordering Information - TTP-VARIO AERIAL Total Prosthesis
ITEM REF
TTP-VARIO AERIAL Total Prosthesis Adjustable Length
3.0 - 7.0 mm (in 0.25 mm increments)
1004 010

 

The TTP-VARIO Total Prosthesis is compatible with the OMEGA CONNECTOR

 

Material: Pure Titanium (ASTM F67 Medical Grade)

Shaft Diameter: 0.2 mm

 

Prescription Devices Caution:

United States Federal Law restricts devices identified herein to sale by or on the order of a physician.
These pages do not replace nor do they set forth the complete contents of the “In­structions for Use” for the products shown here, and are not substitute for reviewing and understanding that important information. Therefore, before using any products included here, please review the entire contents of the respective “Instructions for Use” information.
Because of registration the product is currently not available in all countries.

Produktvideo & weitere Impressionen

Literature

Synchronous ossiculoplasty with titanium prosthesis during canal wall down surgery for advanced cholesteatoma: anatomical and hearing outcomes

Iseri M., Ustundag E., Ulubil S. A., Ozturk M., Bircan O.

Published: The Journal of Laryngology & Otology (2012), 126, 131–135.

 

 

Literature

Preliminary ossiculoplasty results using the kurz® titanium middle ear implants

Heylbroeck Ph., De Vel E., Dhooge I.

Published: 24th Politzer Meeting 31 August – 4 September 2003, Amsterdam, the Netherlands: Selected abstracts

Literature

Impact of Prosthesis Length on Tympanic Membrane's
and Annular Ligament's Stiffness and the Resulting Middle Ear
Sound Transmission

Neudert M., Bornitz M., Lasurashvili N., Schmidt U.,
Beleites Th., Zahnert Th.

Published: Otology & Neurotology: October 2016, Volume 37

 

 

 

DMS 0003994_Rev.01